J Am Dent Assoc, Vol 138, No 10, 1360-1365.
© 2007 American Dental Association

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Herd, S.
Right arrow Articles by Ofner, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Herd, S.
Right arrow Articles by Ofner, S.
Related Collections
Right arrow Dental Equipment/Instruments

RESEARCH

JADA Continuing Education

The in vivo contamination of air-driven low-speed handpieces with prophylaxis angles



Sarah Herd, DDS, MSD, Judith Chin, DDS, MS, Charles John Palenik, MS, MBA, PhD and Susan Ofner, MS


   ABSTRACT
 TOP
 ABSTRACT
 SUBJECTS, MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 CONCLUSION
 REFERENCES
 
Background. The authors conducted an in vivo study to determine if low-speed handpiece motors can become contaminated with oral flora when used with prophylaxis angles.

Methods. This crossover study involved 20 subjects, two types of handpieces and three prophylaxis angles. The authors used each handpiece/prophylaxis angle system to polish teeth. They then collected samples, spiral-plated the specimens and incubated them at 37°C anaerobically and aerobically (with 5 percent carbon dioxide). After incubation, the authors examined the plates for the presence of bacterial colonies.

Results. At least 75 percent of the handpiece/prophylaxis angle systems used on the 20 subjects had bacterial contamination for at least one cultured area. Of the 420 specimens, 258 (61.4 percent) produced bacterial growth. Contamination varied from zero to 6,300 colony-forming units per milliliter.

Conclusions. These data suggest that the internal surfaces of low-speed handpieces can become microbially contaminated during use with prophylaxis angles.

Clinical Implications. Unless low-speed handpieces are sterilized properly after each use, they pose a risk for crossinfection.

Key Words: Infection control; handpiece; prophylaxis angle; contamination; sterilization

Abbreviations: ETSA: Enriched trypticase soy agar • PBS: Phosphate-buffered saline

According to the Centers for Disease Control and Prevention, practitioners should "clean and heat-sterilize hand-pieces and other intraoral instruments that can be removed from the air and waterlines of dental units between patients."1 Studies of high-speed handpieces using dye expulsion have confirmed the potential for retraction of oral fluids into internal compartments of the device.2,3 Studies using laboratory models also indicate the possibility for retention of viral DNA and viable viruses inside both high-speed handpieces and prophylaxis angles.4,5 The potential for contamination of the internal surface of other devices (for example, low-speed handpieces) has not been studied clinically.

The standard of care is to heat-sterilize high-speed dental hand-pieces after each use, as studies have demonstrated the potential for internal contamination during use that could lead to cross-contamination.213 The justification for heat-sterilizing low-speed hand-piece systems is less clear. Compressed air is needed to operate air-driven low-speed handpieces, and it must be reduced or allowed to escape to prevent excessive heat buildup. All disposable and reusable prophylaxis angles have a vent that helps reduce or eliminate excessive heat buildup, and it helps prevent the handpiece from being a sealed system, which may contribute to internal contamination of the handpiece. If bacterial contaminants escape through the vent but remain within the handpiece, this could lead to subsequent cross-contamination unless the low-speed handpiece is heat-sterilized between each use. There is, however, only preliminary information concerning the internal contamination of low-speed handpiece systems during use.14,15

A recent in vitro study demonstrated microbial movement into and out of two types of low-speed handpieces attached to eight different types of disposable or reusable prophylaxis angles.15 Angles were operated in solution of Geobacillus stearothermophilus, and then inner surfaces were sampled. The authors tested outward movement by inoculating spores onto the gears of sterilized handpiece motors. Then they submerged the prophylaxis angles and handpieces in beakers of sterile phosphate-buffered saline (PBS) and operated them. They then collected specimens of PBS from the beakers, as well as specimens from the inside of the angle and nose cone, the gears of the nose cone and motor, to evaluate contamination. Results indicated that the spores traveled into the motor 20 percent of the time. When the motor gears were contaminated, the test bacterium traveled beyond the prophylaxis cup, out of the prophylaxis angle and into the PBS solution in 75 of 160 cases (47 percent). This study indicates that internal contamination of low-speed handpiece motors can occur when used with prophylaxis angles. Movement of bacteria, both inward and outward, was shown for all 14 combinations tested.

These results suggest a need for an in vivo study examining the same hypothesis. We conducted a study to investigate whether low-speed handpieces with prophylaxis angles can become contaminated with oral flora during clinical use. Our hypothesis was that microbes could enter the prophylaxis angle/handpiece system at the prophylaxis angle’s tip and travel to the gears of the air-driven handpiece motor.


   SUBJECTS, MATERIALS AND METHODS
 TOP
 ABSTRACT
 SUBJECTS, MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 CONCLUSION
 REFERENCES
 
Subjects. The Indiana University-Purdue University of Indianapolis Institutional Review Board (approval number 0605–74) reviewed and approved our study protocol. We recruited 20 healthy adult subjects (average age 39.8 years; range, 22–59 years). Our inclusion criteria were that subjects have a minimum of 20 teeth, have good dental health, have a history of a prophylaxis within the last two years and not have a history of untreated periodontal disease. Our exclusion criteria included the subjects’ having any significant medical condition or systemic disease requiring them to undergo subacute bacterial endocarditis prophylaxis or being unable to participate in six visits. Subjects provided written informed consent before participation.

Study design. The subjects participated in a randomized, three-by-two crossover study. They underwent a full-mouth tooth polishing performed by a single investigator (S.H.) once a week for six weeks. Each subject received a full-mouth polishing with two test handpieces connected to each of three prophylaxis angles. We randomized polishing time to three-, four- and five-minute intervals. We then aseptically disassembled the handpieces and took specimens for microbial analysis.

Low-speed handpieces. We used two hand-pieces. One was the Prophy Star (Star Dental, Lancaster, Pa.) single-piece low-speed handpiece. The other was the Titan 3 (Star Dental) two-piece low-speed handpiece. Both handpieces were connected to a dental unit (Excellence series, A-Dec, Newberg, Ore.). We sterilized the hand-pieces in a steam autoclave for 30 minutes at 121°C before use.

Prophylaxis angles. We tested three prophylaxis angles. Two were disposable, and one was reusable. The disposable angles were Nupro revolv (the firm style) (Dentsply, York, Pa.) and Original Green With Regular White Cup (Denticator, Earth City, Mo.). The reusable metal prophylaxis angle was the TS2 Prophy Angle (metal) with Traditional Web Prophy Cup Screw Type Firm White (Young Dental, Carpentersville, Ill.). We sterilized the reusable prophylaxis angle in the same manner as we sterilized the low-speed handpieces. Each prophylaxis angle and its disposable prophylaxis cup came from the same lot. We selected the best- and worst-performing prophylaxis angles from our previous in vitro study for this study.15

Negative controls. We tested both handpieces with each of the three prophylaxis angles attached two times for sterility. We assembled the sterilized handpieces and angles as described above, operated them for five minutes using the same dental units as we used with the subjects and then analyzed them for contamination. We did not use these negative control handpiece/prophylaxis angle systems intraorally, but we ran their motors to show that the ambient air driving the hand-piece motors did not contain microorganisms and to assess the aseptic nature of the procedure used.

Clinical protocol. We assembled the hand-pieces and angles in a biological safety cabinet (Class II, Type A/B3, NuAire, Plymouth, Minn.) using aseptic techniques. We removed the sterilized handpieces from their sterilization pouches and covered them with disposable plastic sleeves (Low Speed Handpiece Sleeve, Denticator). We attached one prophylaxis angle with cup to the handpiece and extended the sleeve to cover the angle. We used the handpiece/prophy angle systems to polish subjects’ teeth for a randomized length of time (three, four or five minutes), using a fine orange polishing paste that did not contain fluoride. Subjects returned at weekly intervals for polishing with another handpiece/prophylaxis angle system until all six combinations were tested.

Microbial analysis. After we clinically tested the handpieces, we disassembled the components aseptically. We collected samples from the two-unit handpiece/prophylaxis angle system from four areas: the internal shaft of the prophylaxis angle where the angle attached to the nose cone, the internal shaft on the superior end of the nose cone, the gears on the inferior end of the nose cone and the gears of the motor (Figure 1Go). We collected samples from the single-piece handpiece/prophylaxis angle system from three areas: the internal shaft of the prophylaxis angle where the angle attached to the nose cone, the internal shaft on the superior end of the nose cone and the gears of the motor. Owing to the design of the single-piece handpiece, which needs to be disassembled completely to access the motor gears, we did not collect samples from the gears on the interior end of the nose cone. We repeated these sampling steps weekly until we tested each type of handpiece/prophylaxis angle system.


Figure 1
View larger version (41K):
[in this window]
[in a new window]

 
Figure 1. The four areas from which specimens were collected from the low-speed handpiece/prophylaxis angle system. A. Example of the inside of the prophylaxis angle. B. Example of the inside of the nose cone. C. Example of gears of the nose cone. D. Example of gears of the motor.

 
We used sterile cotton-tipped applicators and sterile extracoarse paper points to collect samples from the handpiece/prophylaxis angle systems. We swiped the entire area of each site three times for five seconds. We then placed the swabs and paper points individually into 2 milliliters of PBS (0.01 molar, pH 7.2) and mixed them vigorously for 30 seconds, after which we spiral-plated 0.05 mL onto enriched trypticase soy agar (ETSA) plates (BD, Franklin Lakes, N.J.). ETSA is a nonselective medium that supports the growth of many types of microorganisms.

We incubated the ETSA plates anaerobically (GasPak Plus, Disposable Hydrogen + Carbon Dioxide Generator Envelope, BD) at 37°C for seven days and aerobically supplemented them with carbon dioxide (CO2 Incubator, ESPEC, Osaka, Japan). After incubation, we analyzed the ETSA plates for the presence of bacterial colonies. Contamination levels were expressed as colony-forming units per milliliter (CFU/mL).

Each subject contributed three specimens a week when we used a single-piece handpiece and four specimens a week when we used a two-piece handpiece. We treated a single subject with three single-piece handpiece/prophylaxis angle systems for a total of nine specimens. We treated a single subject with three two-piece handpiece/prophylaxis angle systems for a total of 12 specimens. Therefore, the total number of specimens taken from a single subject was 21, and the total number of specimens taken from all subjects was 420.

Data analysis. We defined contamination as a bacterial count greater than zero. We calculated the percentage of samples with contamination for each handpiece/prophylaxis angle system, sampling area and culturing method type. We used separate logistic models fit using generalized estimating equation methodology to compare the overall percentage of contamination between two-piece and single-piece handpieces and the overall percentage of contamination between aerobic and anaerobic cultures. We calculated the total bacteria count for each subject by adding counts across areas and bacteria types (anaerobic and aerobic) for each handpiece/prophylaxis angle system. We compared total counts between handpiece/prophylaxis angle systems using a linear model with parameters for handpiece, prophylaxis angle, handpiece-by-prophylaxis angle interaction and random subject effect. We adjusted comparisons of mean counts between the six handpiece/prophylaxis angle systems by using a Tukey-Kramer adjustment. We compared the mean total count between operation times by using an analysis of variance model for each handpiece/prophylaxis angle system. We used a software package (SAS Version 9.1, SAS, Cary, N.C.) for analyses.


   RESULTS
 TOP
 ABSTRACT
 SUBJECTS, MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 CONCLUSION
 REFERENCES
 
Twenty subjects completed the study, and we collected and cultured 420 specimens. No serious adverse events occurred that were related to the study or otherwise. For each of the six hand-pieces/prophylaxis angle systems, at least 75 percent of the handpiece/prophylaxis angle systems used on the 20 subjects had bacterial contamination for at least one cultured area (Table 1Go and Figure 2Go). Contamination ranged from 75 to 100 percent; the single-piece handpiece with the Nupro revolv prophylaxis angle had the smallest percentage of contamination (75 percent). Of the 420 sites sampled, 258 (61.4 percent) produced bacterial growth. Contamination varied from zero to 6,300 CFU/mL per specimen (0–4,500 CFU/mL for anaerobic bacteria: 0–5,000 CFU/mL for aerobic bacteria).


View this table:
[in this window]
[in a new window]

 
TABLE 1 Percentage of bacterial contamination, by handpiece and prophylaxis angle.

 

Figure 2
View larger version (38K):
[in this window]
[in a new window]

 
Figure 2. Percentage contamination by either bacteria type by handpiece and prophylaxis angle. The prophylaxis angles’ manufacturers are as follows: Original Green With Regular White Cup (Denticator, Earth City, Mo.), TS2 Prophy Angle (metal) (Young Dental, Carpentersville, Ill.) and Nupro revolv (firm) (Dentsply, York, Pa.).

 
When we compared the overall contamination of two-piece handpieces with that of single-piece handpieces, we found no significant difference (Table 2Go). There was no significant difference between the percentage of samples contaminated with aerobic bacteria and the percentage of specimens contaminated with anaerobic bacteria. When we compared the percentage of specimens in each handpiece area contaminated by either anaerobic or aerobic bacteria, we found no significant difference between the areas in two-piece or single-piece handpieces. We found no significant difference in mean total bacteria count (CFU/mL) between any of the handpiece/prophylaxis angle systems (Table 3Go). We polished each subject’s teeth for an average of four minutes, but the polishing times were randomized to three, four or five minutes at each visit. There was no significant difference in the amount of mean bacterial count (CFU/mL) between the randomized operation times (Table 3Go).


View this table:
[in this window]
[in a new window]

 
TABLE 2 Comparison of contamination by handpiece and bacteria type.

 

View this table:
[in this window]
[in a new window]

 
TABLE 3 Mean total bacterial count (CFU/mL*) by operation time, handpiece and prophylaxis angle.

 
The tests of the negative controls did not result in any bacterial growth on the ETSA plates. We found no bacterial contamination after running the sterile handpieces and prophylaxis angles for five minutes. The lack of bacterial growth confirms that we performed the assembly and disassembly of the handpieces/prophylaxis angle systems and culturing techniques in an aseptic manner.


   DISCUSSION
 TOP
 ABSTRACT
 SUBJECTS, MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 CONCLUSION
 REFERENCES
 
The results of our in vivo study support those of an in vitro study examining the same hypothesis.11 In our study, we modeled the operation of polishing teeth in a healthy patient using a handpiece/prophylaxis angle system in an oral environment. In our study, polishing resulted in the movement of microbes to an internal component of the handpiece 75 to 100 percent of the time.

There are limitations to any clinical trial. In our study, we were looking for the presence or absence of any bacteria to justify sterilization of low-speed handpieces. We used only two kinds of handpieces and three types of prophylaxis angles. We did not determine the microbial species in the specimens nor did we know the pathogenic potential for the kinds of bacterial cultured. Instead, we were assessing the presence of oral microflora only. Since the control specimens were negative, we know that the bacterial growth resulted from the subjects. We found higher levels of contamination in this in vivo study than we found in our in vitro study.15 In the in vitro study, we collected specimens of only one type of organism. It is possible the nonselective manner of incubation in this in vivo study resulted in more bacterial growth from the variety of oral microflora that are present at any time. Additionally, the subjects were an inconsistent and uncontrollable source of oral microflora.

These results support our hypothesis that oral bacterial contamination would be found on internal surfaces of handpieces after use. We attempted to simulate typical patient conditions. Low-speed handpieces are used multiple times during a typical day to polish teeth after a dental prophylaxis or after periodontal maintenance treatment. If the internal surfaces of low-speed handpieces become contaminated with oral bacteria, there is a possibility that these organisms can be transmitted to the next patient with whom the handpiece is used. Considering the number of times a handpiece might be used in a day, the potential for cross-contamination is substantial if these devices are not heat-sterilized after each use. In vitro studies have shown that bacterial microbes can be expelled from the low-speed handpiece.14,15 In light of a recent report of patient-to-patient transmission of hepatitis B in the dental office,16 it is possible that viral microorganisms also may be transmitted in this manner.

It also is possible that HIV and other blood-borne pathogens might migrate into the hand-pieces. The procedures we followed in our study did not induce bleeding, but bloodborne pathogens can be present in saliva without blood being visible. We did not test specifically for the presence of bloodborne pathogens in the handpieces.


   CONCLUSION
 TOP
 ABSTRACT
 SUBJECTS, MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 CONCLUSION
 REFERENCES
 
We found that bacterial contamination levels were similar for each of the six handpiece/prophylaxis angle combinations. The results of our in vivo study indicate that the internal surfaces of low-speed handpiece motors can become contaminated with oral bacteria after being used to polish teeth. We recommend that low-speed handpieces be heat-sterilized after each use.


   FOOTNOTES
 

Dr. Herd is in private practice in Ft. Wayne, Ind. When this article was written, she was a graduate student, Department of Periodontics, Indiana University School of Dentistry, Indianapolis.


Dr. Chin is an associate professor, Department of Pediatric Dentistry, Indiana University School of Dentistry, 1121 West Michigan St., Indianapolis, Ind. 46202, e-mail "jchin{at}iupui.edu". Address reprint requests to Dr. Chin.


Dr. Palenik is the director, Department of Infection Control Research and Services, Indiana University School of Dentistry, Indianapolis.


Ms. Ofner is a biostatistician, Department of Biostatistics, Indiana University School of Medicine, Indianapolis.


This study was supported, in part, by a grant to Dr. Herd from the Delta Dental Master Thesis Award Program.


   REFERENCES
 TOP
 ABSTRACT
 SUBJECTS, MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 CONCLUSION
 REFERENCES
 

  1. Kohn WG, Collins AS, Cleveland JL, et al. Guidelines for infection control in dental health-care settings: 2003. MMWR Recomm Rep 2003;52(RR17):1–61.[Medline]

  2. Crawford JJ, Broderius C. Control of cross-infection risks in the dental operatory: prevention of water retraction by bur cooling spray systems. JADA 1988;116(6):685–7.[Abstract]

  3. Lewis DL, Boe RK. Cross-infection risks associated with current procedures for using high-speed dental handpieces. J Clin Microbiol 1992;30(2):401–6.[Abstract/Free Full Text]

  4. Lewis DL, Arens M, Appleton SS, et al. Cross-contamination potential with dental equipment. Lancet 1992;340(8330):1252–4.[Medline]

  5. Bagga BS, Murphy RA, Anderson AW, Punwani I. Contamination of dental unit cooling water with oral microorganisms and its prevention. JADA 1984;109(5):712–6.[Abstract]

  6. Epstein JB, Rea G, Sibau L, Sherlock CH. Rotary dental instruments and the potential risk of transmission of infection: herpes simplex virus. JADA 1993;124(12):55–9.[Abstract]

  7. Epstein JB, Rea G, Sibau L, Sherlock CH, Le ND. Assessing viral retention and elimination in rotary dental instruments. JADA 1995: 126(1):87–92.[Abstract/Free Full Text]

  8. Cottone JA, Molinari JA. State-of-the-art infection control in dentistry. JADA 1991;122(8):33–41.[Abstract]

  9. Molinari JA. Dental handpiece sterilization: historical and technological advances. Compend Contin Educ Dent 1998;19(7):724–7.[Medline]

  10. Sanchez E, Macdonald G. Decontaminating dental instruments: testing the effectiveness of selected methods. JADA 1995;126(3): 359–66.[Abstract/Free Full Text]

  11. Saunders MJ, Cottone JA. What is good enough? Sterilization and its discontents. Tex Dent J 1987;104(9):20–9.[Medline]

  12. Underhill TE, Terezhalmy GT, Cottone JA. Prevention of cross-infections in the dental environment. Compend Contin Educ Dent 1986;7(4):260–4.[Medline]

  13. Molinari JA, Palenik CJ. Environmental surface infection control, 2003. Compend Contin Educ Dent 2004;25(1 supplement):30–7.[Medline]

  14. Waskow JR, Miller CH, Palenik CJ. Justification for sterilizing slow-speed handpiece motors between patients (abstract 3182). J Dent Res 1996;75(special issue):415.

  15. Chin JR, Miller CH, Palenik CJ. Internal contamination of air-driven low-speed handpieces and attached prophy angles. JADA 2006;137(9):1275–80.[Abstract/Free Full Text]

  16. Redd JT, Baumbach J, Kohn W, Nainan O, Khristova M, Williams I. Patient-to-patient transmission of hepatitis B virus associated with oral surgery (published correction appears in J Infect Dis 2007;195[12]:1874.). J Infect Dis 2007;195(9):1311–4.[Medline]





This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Herd, S.
Right arrow Articles by Ofner, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Herd, S.
Right arrow Articles by Ofner, S.
Related Collections
Right arrow Dental Equipment/Instruments


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS